Command
Sequence
CAL:TARE @<
ch_list
>
Correct channel offsets.
CAL:TARE?
Return the success flag from the CAL:TARE
operation.
CAL:STOR TARE
Optional depending on necessity of long
term storage.
CALibration:TARE?
CALibration:TARE?
returns a value to indicate the success of the last
CALibration:TARE
operation.
CALibration:TARE?
returns the value only after the
CALibration:TARE
operation is complete.
•
Returned Value:
Value
Meaning
Further Action
0
Cal OK
None
-1
Cal Error
Query the Error Queue
(SYSTem:ERRor?)
See Error Messages in Appendix B.
Also run *TST?.
-2
No results available
Perform CALibration:TARE
before CALibration:TARE?.
The C-SCPI type for this returned value is int16.
•
Executing
CALibration:TARE
sets the Calibrating bit (bit 0) in Operation Status
Group. Executing
CALibration:TARE?
resets the bit.
•
Related Commands:
CALibration:STORe TARE
,
DIAGnostic:CALibration:TARE[:OTDetect][:MODE]
Command
Sequence
CAL:TARE @<
ch_list
>
Correct channel offsets.
CAL:TARE?
Return the success flag from the CAL:TARE
operation.
CAL:STOR TARE
Optional depending on necessity of long
term storage.
CALibration
Chapter 5
HP E1313/E1413 Command Reference 183