Starting
ARM event, either hardware or software
Acquisition starts
Stopping
Stops when Trigger Count is reached unless count is INF, then use ABORt
Note
TRIGger:COUNt
must be greater than or equal to
CALCulate:AVERage:COUNt
in order to complete the average calculation.
CALCulate:AVERage ON enables
averaging for all channels on HP E1413.
TRIGger:COUNt 8 (must be greater than
or equal to CALC:AVER:COUNt).
In this case, TRIG:COUNT is a multiple of 4.
Every 4 scans will place averaged channel
readings into the FIFO buffer.
CALC:AVER:COUNt 4 sets this value.
SENSe:DATA:FIFO:ALL? to
read
averaged measurements.
Figure 4-7. Averaging Channel Measurements
132 Understanding the HP E1313/E1413
Chapter 4