The next two examples measure type K thermocouples. One example uses a
single Terminal Panel and the other uses two Terminal Panels.
Example Using a Single Terminal Panel:
SENS:REF THER,5000,1,(@100,101,102)
Measures reference temperature measurements on channels 100 to 102.
SENS:FUNC:TEMP TC,K,.06,(@103:131)
Defines channels for temperature measurements.
ROUT:SEQ:DEF (@100,103:107,101,108:123,102,124:131)
Defines the scan list.
Example Using Two Terminal Panels:
SENS:REF THER,5000,1,(@100,101,102,132,133,134)
Measures reference temperature measurements on channels 100 to 102.
SENS:FUNC:TEMP TC,K,.06,(@103:131,135:163)
Defines channels for temperature measurements.
ROUT:SEQ:DEF (@100,103:107,101,108:123,102,
124:131,132,135:139,133,140:155,134,156:163)
Defines the scan list.
Note that each reference channel (100, 101, 102, 132, 133, 134)
immediately precedes its associated measurement channels in the scan list.
Chapter 3
Using the HP E1313/E1413 99
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com