OUTPut:CURRent[:STATe]
OUTPut:CURRent[:STATe] <
enable
>,(@<
ch_list
>)
enables or disables current
output on channels specified in
<
ch_list
>
.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
enable
>
numeric (uint16)
1 | 0 | ON | OFF
none
<
ch_list
>
channel list (string)
100 - 163
none
Comments
•
OUTPut:CURRent:STATe
does not affect a channel’s amplitude setting. A
channel that has been disabled, when re-enabled sources the same current set by
the previous
OUTPut:CURRent:AMPLitude
command.
•
OUTPut:CURRent:STATe
is most commonly used to turn off excitation current to
four-wire resistance (and resistance temperature device) circuits during execution
of
CALibration:TARE
for those channels.
•
This command will cause Error +3000, “ illegal while initiated” if trigger is initiated.
•
Related Commands:
OUTPut:CURRent:AMPLitude
,
CALibration:TARE
•
*RST Condition:
OUTPut:CURRent:STATe OFF
(all channels)
Usage
OUTP:CURR:STAT OFF,(@100,108)
Turn off current source channels 0 and 8.
OUTPut:CURRent[:STATe]?
OUTPut:CURRent[:STATe]? (@<
channel
>)
returns the state of the Current
Source SCP channel specified by
<
channel
>
.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
channel
>
channel list (string)
100 - 163
none
Comments
•
<
channel
>
must specify a single channel only.
•
Returned Value: Returns
1
for enabled,
0
for disabled. The C-SCPI type
returned is uint16.
•
Related Commands:
OUTPut:CURRent[:STATe]
,
OUTPut:CURRent:AMPLitude
Usage
OUTP:CURR:STAT? (@108)
Query for state of Current SCP channel 8.
Execute enter statement here
Enter query value, either 1 or 0.
OUTPut