ASSIGN <
channel
> <
conversion
> <
compensation
>
0010
16
ASSIGN for Standard
EU Conversions
(See the next discussion for Custom EU Conversions).
ASSIGN
loads the
specified channel’s linear table area with the correct values for the indicated
conversion. If
<
conversion
>
is from 1 through 1516, it indicates a standard
linearization, such as an RTD or a type K thermocouple. Thermocouple
conversions incorporate reference junction compensation. Some
conversions are specifically for transducers measuring a reference junction.
If
<
conversion
>
is zero, no conversion is performed and the channel returns
voltage. The supported standard conversions are:
Code
EU Conversion
Use Range:
0
16
Volts
0 - 4 | FFFF
16
1
16
Reserved straight-line conversion
2
16
Ohms, 31
µ
A
FFFF
16
3
16
Ohms, 488
µ
A
FFFF
16
4
16
Reference, Thermistor, 5K, 122
µ
A
3
5
16
Reference, RTD, 85, 121
µ
A
0
6
16
Reference, RTD, 92, 121
µ
A
0
7
16
Reference, Custom
0
8
16
Thermocouple, Type J
0
9
16
Thermocouple, Type K
0
A
16
Thermocouple, Type T
0
B
16
Thermocouple, Type E (high accuracy)
0
C
16
Thermocouple, Type E (extended range)
1
D
16
Thermocouple, Type R
0
E
16
Thermocouple, Type S
0
F
16
Thermocouple, Custom, no compensation
0
10
16
Thermistor, 2.25K, 488
µ
A
FFFF
16
11
16
Thermistor, 5K, 31
µ
A
FFFF
16
12
16
Thermistor, 10K, 31
µ
A
FFFF
16
13
16
RTD, 85, 488
µ
A
0
14
16
RTD, 92, 488
µ
A
0
15
16
16
16
- 1F
16
Thermocouple, Type N
Reserved (not used)