INITiate:IMMediate
time if the filter is
ON
and any channel specifies autorange.
<
range
>
may be specified in millivollts (mv).
•
If you are using amplifier SCPs, you should set them first and keep their settings
in mind when specifying a range setting. For instance, if your expected signal
voltage is to be approximately .1 VDC and the amplifier SCP for that channel has
a gain of 8, you must set
<
range
>
no lower than 1 VDC or an input out-of-range
condition will exist.
•
The
*CAL?
command calibrates temperature channels based on Sense Amplifier
SCP setup at the time of execution. If SCP settings are changed, those channels
are no longer calibrated.
*CAL?
must be executed again.
•
Related Commands:
DIAGnostic:CUSTom:PIECewise
,
[SENSe:]FUNCtion:TEMPerature
,
[SENSe:]FUNCtion:CUSTom:TCouple
,
*CAL?
•
*RST Condition: All custom EU tables are erased.
Usage
Program must put table constants into array table_block.
DIAG:CUST:PIEC 1,table_block,(@108)
Send characterized reference transducer
table for use by channel 8.
SENS:FUNC:CUST:REF .25,(@108)
Link custom reference temperature EU with
channel 8.
Include this channel in a scan list with thermocouple channels (REF channel first).
INITiate then TRIGger module.
[SENSe:]FUNCtion:CUSTom:TCouple
[SENSe:]FUNCtion:CUSTom:TCouple <
type
>,[<
range
>,](@<
ch_list
>)
links
channels with the custom Engineering Unit Conversion table loaded with the
DIAGnostic:CUSTom:PIECewise
command. The table is assumed to be for a
thermocouple and the
<
type
>
parameter will specify the built-in compensation
voltage table to be used for reference junction temperature compensation.
SENSe:FUNCtion:CUSTom:TCouple
allows you to use an EU table that is custom
matched to thermocouple wire you have characterized. Contact your
Hewlett-Packard System Engineer for more information on Custom Engineering
Unit Conversion for your application.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
type
>
discrete (string)
E | J | K | N | R | S | T
none
<
range
>
numeric (float32)
.0625 | .25 | 1 | 4 | 16
(See Comments)
VDC
<
ch_list
>
channel list (string)
100 - 163
none
Comments
•
See “ Linking Channels to EU Conversion” in Chapter 3 for more information.
[SENSe:]
238 HP E1313/E1413 Command Reference
Chapter 5