Command Quick Reference
The following tables summarize SCPI and IEEE-488.2 Common (*) Commands for
the HP E1313/E1413 64-Channel High-Speed A/D.
SCPI Command Quick Reference
Command
Description
ABORt
Stops scanning immediately and sets trigger system to idle state (scan
lists are unaffected).
ARM
[:IMMediate]
Arm if ARM:SOURce is BUS or HOLD (software ARM).
:SOURce BUS | EXT | HOLD | IMM | SCP | TTLTrg<n>
Specify the source of Trigger Timer ARM.
:SOURce?
Return current ARM source.
CALCulate
:AVERage
[:STATe] ON | OFF
Enables/disables channel measurement averaging.
[:STATe]?
Returns the state of channel averaging.
:COUNt <n>
Sets the number of measurements averaged to produce a stored reading.
:COUNt?
Returns the current setting of measurements per reading.
:CLIMits
:FAIL
[:CUMulative]?
Returns composite limit test status since module was INITiated.
:CURRent?
Returns composite limit test status for last completed scan list.
:FLIMits
[:CHANnels]
[:CUMulative]?
Returns all channel’s limit test status since module was INITiated.
:CURRent?
Returns all channel’s limit test status for last completed scan list.
:POINts
[:CUMulative]?
Returns count of channels exceeding limit tests since module INITiated.
:CURRent?
Returns count of channels exceeding limit tests for last completed scan.
:LIMit
[:STATe] ON | OFF,(@<ch_list>)
Enables/disables all limit testing.
[:STATe]? (@<channel>)
Returns state of limit testing.
:FAIL
[:CUMulative]? (@<channel>)
Returns limit test status for channel since module was INITiated.
:CURRent? (@<channel>)
Returns limit test status for channel for last completed scan list.
:LOWer
[:STATe] ON | OFF,(@<ch_list>)
Enables/disables lower limit testing.
[:STATe]? (@<channel>)
Returns state of lower limit testing.
:DATA <lower_lim>,(@<ch_list>)
Sets lower limit for specified channels.
:DATA? (@<channel>)
Returns lower limit setting for specified channel.
:UPPer
[:STATe] ON | OFF,(@<ch_list>)
Enables/disables upper limit testing.
[:STATe]? (@<channel>)
Returns state of upper limit testing.
:DATA <upper_lim>,(@<ch_list>)
Sets upper limit for specified channels.
:DATA? (@<channel>)
Returns upper limit setting for specified channel.
CALibration
:CONFigure
Prepare to measure on-board references with an external multimeter.
:RESistance
Configure to measure reference resistor.
:VOLTage <range>, ZERO | FSCale
Configure to measure reference voltage range at zero or full scale.
:SETup
Performs Channel Calibration procedure.
:SETup?
Returns state of CALibration:SETup operation (returns error codes or 0
for OK).
:STORe ADC | TARE
Store cal constants to Flash RAM for either A/D calibration or those
generated by the CALibration:TARE command.
Command Quick Reference
284 HP E1313/E1413 Command Reference
Chapter 5