CALibration:TARE:RESet
CALibration:TARE:RESet
resets the tare calibration constants to zero for all
64 channels. Executing
CALibration:TARE:RESet
affects the tare cal constants in
RAM only. To reset the tare cal constants in Flash Memory, execute
CALibration:TARE:RESet
and then execute
CALibration:STORe TARE
.
Command
Sequence
CAL:TARE:RES
Reset channel offsets.
CAL:STOR TARE
Optional if necessary to reset tare cal
constants in Flash Memory.
CALibration:VALue:RESistance
CALibration:VALue:RESistance <
ref_ohms
>
sends the just-measured value of
the on-board reference resistor to the module for A/D calibration.
Parameters
Parameter
Name
Parameter
Type
Range of
Value
Default
Units
<
ref_ohms
>
numeric (float32)
7,500
±
5%
ohms
Comments
•
<
ref_ohms
>
must be within 5% of the nominal reference resistor value (7,500 Ohms)
and may be specified in Kohm (kohm).
•
A four-wire measurement of the resistor can be made with an external multimeter
connected to the H Cal, L Cal, H ohm, and L ohm terminals on the terminal
module, or the V H, V L,
Ω
Ω
H, and
Ω
Ω
L terminals on the Cal Bus connector.
•
Use the
CALibration:CONFigure:RESistance
command to configure the reference
resistor for measurement at the Cal Bus connector.
•
Use the
CALibration:CONFigure:RESistance
command prior to
CALibration:VALue:RESistance
or Error +3004, “ illegal command” will occur.
•
Related Commands:
CALibration:CONFigure:RESistance
,
CALibration:STORe ADC
Command
Sequence
CAL:CONF:RES
Now measure voltage with external DMM.
CAL:VAL:RES <
measured value
>
Send measured value to module.
CALibration
184 HP E1313/E1413 Command Reference
Chapter 5