Example Command Sequence
Card 1
ROUT:SEQ:DEF LIST<
n
>,(@101,103,105,107)
<n> can be 1 - 4.
ROUT:SCAN LIST<
n
>
SAMP:TIM LIST<
n>,<t
c
>
<t
c
> can be 10
µ
s to 32.768ms.
TRIG:SOUR TIM
TRIG:TIM:PER <
t
p
>
<t
p
> can be .1ms - 6.5536 s.
TRIG:TIM:MODE SYNC
Keep timer running during
INIT:CONT ON.
TRIG:COUN 2
ARM:SOUR <
source
>
<source> can be BUS, EXT,
HOLD, IMM, or TTLTrg0 -
TTLTrg7.
OUTP:TTLT<
n
>:SOUR FTR
1st trig will source TTLTrg.
OUTP:TTLT<
n
>:STAT ON
Enable to drive selected TTLTrg
line. <n> can be 1 through 7.
INIT:CONT ON
Card 2
ROUT:SEQ:DEF LIST<
n
>,(@101,103,105,107)
<n> can be 1 - 4.
ROUT:SCAN LIST<
n
>
SAMP:TIM LIST<
n
>,<
t
c
>
<t
c
> can be 10
µ
s to 32.768ms.
TRIG:SOUR TTLTrg<
n
>
Selects TTLTrg trigger source.
TRIG:COUN <
count
>
<count> can be 1 - 32768 or INF.
INIT:IMM
Starting
ARM event, either hardware or software
Acquisition starts
Stopping Before Trigger Count is Reached
INIT:CONT OFF
To master card.
TRIG:SOUR HOLD
Stops at end of scan list, still
initiated.
TRIG:IMM
One more scan, now idle.
or
ABOR
Stops immediately (must use
ABOR if TRIG:COUNt is INF).
Chapter 4
Understanding the HP E1313/E1413 119
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com