*RST Condition:
SENSe:FUNCtion:VOLTage AUTO,(@100:163)
Usage
FUNC:RES 30ua,(@100,105,107)
Set channels 0, 5, and 7 to convert voltage
to resistance assuming current source set to
30
µ
A use autorange (default).
[SENSe:]FUNCtion:STRain:FBENding
[SENSe:]FUNCtion:STRain:FBPoisson
[SENSe:]FUNCtion:STRain:FPOisson
[SENSe:]FUNCtion:STRain:HBENding
[SENSe:]FUNCtion:STRain:HPOisson
[SENSe:]FUNCtion:STRain[:QUARter]
Note on Syntax: Although the strain function is comprised of six separate SCPI
commands, the only difference between them is the bridge type they specify to the
strain EU conversion algorithm.
[SENSe:]FUNCtion:STRain:<
bridge_type
> [<
range
>,](@<
ch_list
>)
links the strain
EU conversion with the channels specified by
<
ch_list
>
to measure the bridge voltage.
See “ Linking Channels to EU Conversion” in Chapter 3 for more information.
<
bridge_type
>
is not a parameter but is part of the command syntax. The following
table relates the command syntax to bridge type. See the User’s Manual for the
optional Strain SCP for bridge schematics and field wiring information.
Command
Bridge Type
:FBENding
Full Bending Bridge
:FBPoisson
Full Bending Poisson Bridge
:FPOisson
Full Poisson Bridge
:HBENding
Half Bending Bridge
:HPOisson
Half Poisson Bridge
[:QUARter]
Quarter Bridge (default)
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
range
>
numeric (float32)
.0625 | .25 | 1 | 4 | 16
(See Comments)
VDC
<
ch_list
>
channel list (string)
100 - 163
none
[SENSe:]
Chapter 5
HP E1313/E1413 Command Reference 241
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com