Sending the
ABORt
command will immediately send the trigger system back to its
Trigger Idle state and terminate any scan in progress. If
INITiate:CONTinuous
is
ON
, however, it will immediately return to the Wait-for-Trigger state. If
TRIGger:SOURce
is
IMMediate
also, then the module will resume scanning. Use
INITiate:CONTinuous OFF
to stop after the current scan is complete. You can
check the “ Measuring” bit (bit 4) with
STATus:OPERation:CONDition?
to
determine when the scan has completed.
•
Sending
INITiate:IMMediate
while the system is still in the Wait-for-Trigger state
(already
INITiated
) will cause Error -213, “ Init ignored” .
•
Related Commands:
ABORt
,
TRIGger[:IMMediate]
,
TRIGger:SOURce
•
*RST Condition: Trigger system is in the Trigger Idle state.
Usage
INIT:CONT ON
Return to Initiated State after each trigger.
INIT:CONT OFF
Finish scan (if triggered) and return to
Trigger Idle state AFTER the next trigger
event if already armed.
INITiate[:IMMediate]
INITiate[:IMMediate]
changes the trigger system from the Idle state to the
Wait-for-Trigger state. When triggered, one or more (depending on
TRIGger:COUNt
) trigger cycles occur and the instrument returns to the Trigger Idle
state.
ROUTe:SCAN LIST1
through
LIST4
specifies which of four scan lists to
execute.
Comments
•
INITiate:IMMediate
clears the FIFO and Current Value Table.
•
If a trigger event is received before the instrument is initiated, Error -211,
“ Trigger ignored” error is generated.
•
If another trigger event is received before the instrument has completed the
current trigger cycle (measurement scan), the Questionable Data Status bit 9 is set
and Error +3012, “ Trigger too fast” error is generated.
•
Sending
INITiate:IMMediate
while the system is still in the Wait-for-Trigger state
(already
INITiated
) will cause Error -213,"Init ignored".
•
Sending the
ABORt
command will immediately send the trigger system to the
Trigger Idle state and terminate any scan in progress.
•
Related Commands:
ABORt
,
TRIGger
•
*RST Condition: Trigger system is in the Idle state.
Usage
INIT:IMM
Both versions same function.
INITiate:IMMediate
INITiate
Chapter 5
HP E1313/E1413 Command Reference 203