INPut
The
INPut
subsystem controls configuration of programmable input Signal
Conditioning Plug-Ons (SCPs).
Subsystem Syntax
INPut
:FILTer
[:LPASs]
:FREQuency <
cutoff_freq
>,(@<
ch_list
>)
:FREQuency? (@<
channel
>)
[:STATe] <
enable
,(@<
channel
>)
[:STATe]? (@<
channel
>)
:GAIN <
chan_gain
>,(@<
ch_list
>)
:GAIN? (@<
channel
>)
:LOW <
wvolt_type
>,(@<
ch_list
>)
:LOW? (@<
channel
>)
INPut:FILTer[:LPASs]:FREQuency
INPut:FILTer[:LPASs]:FREQuency <
cutoff_freq
>,(@<
ch_list
>)
sets the cutoff
frequency of the filter on the specified channels.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
<
cutoff_freq
>
numeric (float32)
(string)
See Comments below.
MIN | MAX
Hz
<
ch_list
>
channel list (string)
100 - 163
none
Comments
•
<
cutoff_freq
>
may be specified in kiloHertz (khz). A programmable Filter SCP
has a choice of several discrete cutoff frequencies. The cutoff frequency set will
be the one closest to the value specified by cutoff_freq. Refer to Chapter 6 for
specific information on the SCP you are programming.
•
Sending
MAX
for the cutoff_freq selects the SCP’s highest cutoff frequency.
Sending
MIN
for the cutoff_freq selects the SCP’s lowest cutoff frequency.
To disable filtering (the “ pass through” mode), execute the
INPut:FILTer:LPASs:STATe OFF
command.
•
Sending a value greater than the SCP’s highest cutoff frequency or less than the
SCP’s lowest cutoff frequency generates Error -222, “ Data out of range” error.
•
This command will cause Error +3000, “ illegal while initiated” if trigger is initiated.
•
Related Commands:
INPut:FILTer[:LPASs]:FREQuency?
,
INPut:FILTer[:LPASs][:STATe]
•
*RST Condition: Set to
MIN
.