Card 2
ROUT:SEQ:DEF LIST<
n
>,(@101,103,105,107)
<n> can be 1 - 4.
ROUT:SCAN LIST<
n
>
SAMP:TIM LIST<
n>,<t
c
>
<t
c
> can be 10
µ
s to 32.768ms.
TRIG:SOUR IMM
ARM:SOUR <
source
>
<source> can be BUS, EXT,
HOLD, IMM, or TTLTrg0 -
TTLTrg7.
INIT:CONT ON
Starting
ARM event, either hardware or software
Acquisition starts
Stopping
INIT:CONT OFF
To both cards, stops at end of
scan list.
Internal Timer
Based Scans
In this case both cards are set up to be paced by their trigger timers. Card 1 is
also set up to source a trigger on one of the VXIbus TTLTRG lines and card 2
is set up to be triggered by that line. This means that triggering card 1 will start
both cards.
t
c
t
p
t
c
is the sample time between channels.
t
p
is the pacing time between channel lists.
The large arrow is the pacing Trigger Timer. The smaller arrows are sample
triggers generated by the HP E1313/E1413 when stepping through the
channel list. The first large arrow is the arming event.
Channel Number
=Trigger Count
Card 1
Master
Card 2
Slave