Modbus
169
17408
4
clk
Yes
Yes
Request start time
“ Data format
example” (page 174)
HISTORY
(DAY
REPORT)
17412
4
clk
Yes
Yes
Request completion time
“ Data format
example” (page 174)
17416
1
short
Yes
Yes
Request line
Line number
17417
1
short
Yes
Yes
Request element
“ Data definition sheet”
(page 177)
17418
1
short
Yes
No
Unused area
Reading: Constantly 0
Writing: Ignored
17419
1
short
Yes
Yes
Request status
0: No response, 1:
Initialization, 2:
Subsequent data, 3:
Same data
17420
30
day_re
p_dat
Yes
No
Day report history 01
“ Data format
example” (page 174)
17920
4
clk
Yes
Yes
Request start time
“ Data format
example” (page 174)
HISTORY
(ALARM)
17924
4
clk
Yes
Yes
Request completion time
“ Data format
example” (page 174)
17928
1
short
Yes
No
Unused area
Reading: Constantly 0
Writing: Ignored
17929
1
short
Yes
No
Unused area
Reading: Constantly 0
Writing: Ignored
17930
1
short
Yes
No
Unused area
Reading: Constantly 0
Writing: Ignored
17931
1
short
Yes
Yes
Request status
0: No response, 1:
Initialization, 2:
Subsequent data, 3:
Same data
17932
7
memal
m
Yes
No
Alarm history 01
“ Data format
example” (page 174)
18432
4
clk
Yes
Yes
Request start time
“ Data format
example” (page 174)
HISTORY
(CALIBRATI
ON)
18436
4
clk
Yes
Yes
Request completion time
“ Data format
example” (page 174)
18440
1
short
Yes
Yes
Request element
“ Data definition sheet”
(page 177)
18441
1
short
Yes
No
Unused area
Reading: Constantly 0
Writing: Ignored
18442
1
short
Yes
No
Unused area
Reading: Constantly 0
Writing: Ignored
18443
1
short
Yes
Yes
Request status
0: No response, 1:
Initialization, 2:
Subsequent data, 3:
Same data
18444
44
calib_d
ata
Yes
No
Calibration history 01
“ Data format
example” (page 174)
Register
Modbus
address
Length
Data
format
Read Write
Item name
Read/write content
Classification