Function
82
FLOW ALARM SET
Specifies the flow rate alarm value.
Operation procedure
1.
Press [FLOW ALARM SET] in the MEAS. LINE SET screen.
The FLOW ALARM SET screen will be displayed.
2.
Specify the flow rate alarm value.
Item
Setting range
Default value
Description
Hi Limit L1 to L2
0.00 m
3
/d to 9999999 m
3
/d
0.00 m
3
/d
Specify the flow rate alarm value.
Note 1: L2 can be selected only for the two-point meter specification.
Note 2: When these parameters are set to 0.0000, no alarm will be activated.
Main
MENU
SETTING
MEAS. LINE SET
[MENU]
[MEAS. LINE SET]
[SETTING]
Refer to “ To display the [MENU] item screen” (page 62).
FLOW ALARM SET
MEAS. LINE SET
Hi Limit
[FLOW
Hi Limit L1 <Input a value>
L1
ALARM SET]
[ ]