Function
108
COMMUNICATION SET
Specifies the communication settings.
Operation procedure
1.
Press [SYSTEM SET] in the SETTING screen.
The SYSTEM SET 1/2 screen will be displayed.
2.
Press [COMMUNICATION SET].
The COMMUNICATION screen will be displayed.
3.
Specifies the communication settings.
Item
Setting range
Default value
Description
Machine ID
1 to 247
1
Specifies the ID number of the meter.
Baudrate
9600 bps
19200 bps 38400 bps
19200 bps
Allows you to select the communication
speed.
Parity
NONE
ODD
EVEN
NONE
Specifies the communication parity.
[MENU]
Main
Refer to “ To display the [MENU] item screen” (page 62).
[SETTING]
SETTING
MENU
SYSTEM SET 1/2
[COMMUNICATION SET]
SETTING
[SYSTEM
Machine ID
Machine
Baudrate
Parity
Parity
COMMUNICATION
Baudrate
<Input a value>
SET]
ID
[ ]
[ ]
[ ]