
Maintenance
203
6.
Pull down the safety lever and push in the pure water cartridge with the in/out ports
heading forward, then pull up the safety lever.
Note
If the safety lever can not be pulled up, the pure water cartridge may be inserted incompletely, push it
again.
7.
Open the V-3 valve gradually to purge the pure water cartridge with tap water.
8.
Press
【
MENU
】
-
【
MAINTENANCE
】
-
【
SEPARATE ACTION
】
-
【→】
on the screen.
The SEPARATE ACTION 5/5 menu will be opened.
9.
Press
【
Pure water purge
】
to start purging of the pure water cartridge.
Basic purge time ; about 5 minutes
(Depending on the water quality of the tap water)
10.
About 5 minutes later, press STOP to stop purging.
Note
Discard the used pure water cartridge according to the local rules.
11.
Calibration should be done after the cartridge is replaced. Refer to the
【
START
】
for
the calibration procedure.
【
MENU
】
【
MAINTENANCE
】
Main screen
MENU
MAINTENANCE
【
SEPARATE ACTION
】
SEPARATE ACTION 5/5
【→】