Function
126
EXT. ANALOG OUTPUT
Allows you to check the external analog output item and its setting value. The setting value
can be changed.
Operation procedure
1.
Press [EXT. ANALOG OUTPUT] in the CHECK 1/2 screen.
The EXT. DA ADJUST screen will be displayed.
The external analog output item and its setting value can be checked.
2.
To change the setting value, select the item.
3.
Select the setting value and press [DECIDE].
4.
Press [OUTPUT].
Note
After approximately 10 minutes have passed with no button operation occurring on the screen within
the
outlined area shown in the illustration below, output will revert to the setting specified
for ch.
Item
Setting range
Default value
Description
ch00
0.0 mA
4.0 mA
8.0 mA
12.0 mA
16.0 mA
20.0 mA
0.0 mA
Setting of the ch00 analog output
ch01
Setting of the ch01 analog output
ch02
Setting of the ch02 analog output
ch03
Setting of the ch03 analog output
[MENU]
Main
Refer to “ To display the [MENU] item screen” (page 62).
[CHECK]
CHECK 1/2
MENU
CHECK 1/2
[EXT. ANALOG OUTPUT]
EXT. DA ADJUST
<Specify
various
settings>
[ ]