MicroDock II Automatic Test and Calibration Station
9
Introduction
WArNINg
To ensure personal safety, read the Safety Information before using the
MicroDock II Base Station.
The MicroDock II Automatic Test and Calibration Station (“the station”) provides
automated calibration, bump testing and data transfer for the XD and X5
detectors.
A maximum of ten modules can be connected to the station.
NOTe
There is a maximum of six charging modules.
Six charging modules plus four non-charging modules can be connected to the
station.
Safety Information - Read First
Use the station only as specified in this manual. International symbols used on the
station and in this manual are explained in Table 1.
Read the Caution statements on the following pages before using the station.
NOTe
This instrument contains batteries. Do not mix with the solid waste stream.
Spent batteries should be disposed of by a qualified recycler or hazardous
materials handler.
CAUTION
MicroDock ll Base Station
If the station is damaged or parts are missing, contact Honeywell Analytics
»
immediately.
The station must be used only in an area that is free of background gas. Do
»
not use the station in a hazardous area. Failure to adhere to this caution
can lead to fire and/or explosion.
This equipment uses potentially harmful gas for calibrations. The station must
»
be attached to a venting system or be used in a well-ventilated area.
Perform calibrations and bump checks only in a clean atmosphere that is
»
free of background gas.
The maximum recommended exhaust line length is 50 ft. (15.24 m).
»
Ensure that the inlet filter is clean.
»
Ensure that all gas cylinders contain enough gas.
»
Ensure the exhaust line is not connected to a negative pressure system.
»
A demand flow regulator must be used with all gas cylinder connections.
»
Do not expose the station to electrical shock or severe continuous
»
mechanical shock.
The station warranty will be void if the unit is disassembled, adjusted, or
»
serviced by non-Honeywell Analytics personnel.
Do not immerse the station in liquids.
»
Summary of Contents for MicroDock II
Page 1: ...MicroDock II Operating Manual MICRODOCK II AUTOMATIC TEST AND CALIBRATION ...
Page 2: ......
Page 8: ...MicroDock II Automatic Test and Calibration Station 8 ...
Page 38: ...MicroDock II Automatic Test and Calibration Station 38 ...
Page 64: ...MicroDock II Automatic Test and Calibration Station 64 ...
Page 67: ......