MicroDock II Automatic Test and Calibration Station
48
If more than one docking module is queued: If more than one docking
module is queued for a bump check, the first module queued begins the
bump check.
The RUN LED flashes yellow on the remaining docking modules that are
queued for a bump check. When the first module is complete, the next
module in the queue begins the bump check.
From
4.
the docking module, press
BUMP CHECK. The RUN LED lights
yellow.
Depending upon the type of detector, the option to reconfigure is
5.
provided.
If the detector can be reconfigured, the following two screens display.
To reconfigure the detector, press
Yes. The MMC card then transfers the
new configuration data to the detector. If required, refer to
Reconfiguring
the Detector
.
Or
Press
No to continue with the bump check.
NOTe
If a selection is not made within 15 seconds, the station automatically
defaults to the
No
selection and proceeds with the bump check.
The station begins the bump check and displays the
6.
bay number,
•
process being performed,
•
type of detector, and
•
serial number of the detector.
•
NOTe
The X5 beeps and flashes twice to test the sensors when the bump check
begins.
If there is a gas conflict or a selected gas cannot be found, the gas select
7.
screen displays.
Gas conflicts must be resolved to continue the bump check. Refer to
Gas
Conflicts
.
Summary of Contents for MicroDock II
Page 1: ...MicroDock II Operating Manual MICRODOCK II AUTOMATIC TEST AND CALIBRATION ...
Page 2: ......
Page 8: ...MicroDock II Automatic Test and Calibration Station 8 ...
Page 38: ...MicroDock II Automatic Test and Calibration Station 38 ...
Page 64: ...MicroDock II Automatic Test and Calibration Station 64 ...
Page 67: ......