MicroDock II Automatic Test and Calibration Station
45
Accessing the MMC Card During a Test
If an attempt is made to access the MMC card on the station from a PC while a bump
check or calibration is being performed, the following screen displays.
NOTe
Honeywell Analytics recommends that bump checks and calibrations be
completed before attempting to access test results from the PC.
Order of Gases Applied for Bump Checks and Calibrations
WArNINg
It is extremely important that gases are applied in the order that is specified in
this manual.
Failure to adhere to the rules in Table 17 and Table 18 of this manual will
result in incorrect sensor readings and possible personal injury and/or property
damage.
NOTe
Manual application of gas types is available for the X5 detectors only.
To manually apply gases for bump checks and calibrations, complete the
following:
Review
1.
Table 17. Gas Type Application Table
, and
Table 18. Gas
Application Rules
.
Determine the gases that are to be applied.
2.
Using
3.
Table 17
, locate the first gas type (sensor) to be bump checked or
calibrated on the top row.
Within the far left column, locate the next gas type (sensor) to be bumped
4.
or calibrated and refer to the rule number (if applicable).
Using
5.
Table 18
, locate the corresponding rule number. Ensure that the
desired order to apply gases corresponds to the gas application rules.
Continue to reference the table and rules to ensure the gases will be applied
6.
in the correct order.
When the correct order is determined, proceed to the
7.
Bump Check
or
Calibration
section.
Summary of Contents for MicroDock II
Page 1: ...MicroDock II Operating Manual MICRODOCK II AUTOMATIC TEST AND CALIBRATION ...
Page 2: ......
Page 8: ...MicroDock II Automatic Test and Calibration Station 8 ...
Page 38: ...MicroDock II Automatic Test and Calibration Station 38 ...
Page 64: ...MicroDock II Automatic Test and Calibration Station 64 ...
Page 67: ......