MicroDock II Automatic Test and Calibration Station
22
Pump Setup
The pump speed must be set for each new docking module that is connected.
WArNINg
Perform the pump setup procedures in a clean atmosphere only.
Reactivate the station and wait until the normal operating screen displays.
11.
Connect the hose to a flow meter and to inlet 1 (PURGE) on the station.
12.
Allow the flow meter sufficient time to stabilize (approximately 30 minutes)
before using to ensure accurate flow readings.
From the station, access the user options menu.
13.
Press
14.
or
^ to scroll to the Pump Setup option.
Press
15.
sel to activate the pump setup field (: changes to *).
The station immediately begins pumping the ambient air.
The factory default pump speed displays beside Pump Setup. The station
pump setup is measured as a percentage (%) and the flow meter is
measured in ml/min.
NOTe
The station is shipped with the factory default pump speed set to 350
ml/min. (
40-45%
).
Monitor the flow meter until the unit stabilizes (5-20 minutes depending
16.
upon the type of flow meter).
From the station, press
17.
or
^ until the flow meter displays 350 ml/
m.
From the station, press
18.
exit to save the new value and deactivate the
Pump Setup field (* changes to :).
Press
19.
exit again to return to the normal operating screen.
Repeat steps # 1-19 for each docking module that is added. Continue to
20.
step #21 and 22 for the last module added.
After all of the modules have been added, attach the end plate (refer to
21.
Figure 9).
Fasten the bottom cover plate to the space between the two docking modules
22.
using six Philips flat-head screws (refer to Figure 10).
NOTe
When assembling or reassembling parts of the station and docking modules,
refer to the
Table 11 Torque Specifications
.
Summary of Contents for MicroDock II
Page 1: ...MicroDock II Operating Manual MICRODOCK II AUTOMATIC TEST AND CALIBRATION ...
Page 2: ......
Page 8: ...MicroDock II Automatic Test and Calibration Station 8 ...
Page 38: ...MicroDock II Automatic Test and Calibration Station 38 ...
Page 64: ...MicroDock II Automatic Test and Calibration Station 64 ...
Page 67: ......