MicroDock II Automatic Test and Calibration Station
35
If No is selected, the LCD automatically returns to the user options screen. If there
is a problem with the card, the following screen displays.
Press
OK to acknowledge and insert a new card and repeat the procedures.
Press
7.
or
^ to scroll to another user option.
Or
Press
exit to return to the normal operating screen.
For additional information about the MMC card, refer to
Base Station MMC Card
.
Inlet Select
NOTe
Applicable only to custom and multi-gas types.
The Inlet Sel option is used to enable either the automatic (auto) or manual (manu)
function for selecting an inlet.
If the auto option is enabled, the station automatically selects the correct gas inlet
for the bump check or calibration.
If manu is enabled, the applicable test gas must be selected each time a bump/
calibration is performed. If the manu option is enabled, refer to
Order of Gases
Applied for Bump Checks and Calibrations
for important information.
To enable either the auto or manu option, complete the following:
From the user options menu, press
1.
or
^ to scroll to the Inlet Sel
option.
Press
2.
sel to confirm the selection and activate the field (: changes to *).
Press
3.
or
^ to toggle between the auto and manu option.
When the desired option displays, press
4.
sel to confirm the selection and
deactivate the field (* changes back to :).
Press
5.
or
^ to scroll to another user option.
Or
Press
exit to return to the normal operating screen.
Summary of Contents for MicroDock II
Page 1: ...MicroDock II Operating Manual MICRODOCK II AUTOMATIC TEST AND CALIBRATION ...
Page 2: ......
Page 8: ...MicroDock II Automatic Test and Calibration Station 8 ...
Page 38: ...MicroDock II Automatic Test and Calibration Station 38 ...
Page 64: ...MicroDock II Automatic Test and Calibration Station 64 ...
Page 67: ......