MicroDock II Automatic Test and Calibration Station
21
Figure 10. Attaching Back Cover Plate (Back View)
Deactivate the station.
1.
Remove the power cord from the POWER port.
2.
Remove the end plate. There are three Phillips screws on the front and three
3.
Philips screws on the back.
Attach the new docking module.
4.
Ensure the barbed fitting ports mate correctly with the gasket block. Ensure
5.
that the male and female PCB connectors mate correctly.
Initializing the New Docking Module
When a new docking module is connected, it must be initialized (setup to
communicate with the station).
Activate the station.
6.
Simultaneously press and hold
7.
BUMP CHECK on the new docking module
while pressing
(left most button) on the station.
All of the LEDs on the new docking module light. On the station, the
8.
following screen displays to confirm that the new docking module has been
successfully added.
Press
9.
Ok to save the setting.
Deactivate the station.
10.
WArNINg
The station must be deactivated after each module has been added.