MicroDock II Automatic Test and Calibration Station
27
WArNINg
Failed tests can result if the inlets are not setup correctly.
For initial station activation, ensure the inlets are installed correctly as follows:
Inlet 1 is the default connection for ambient air.
•
Inlet 2 is the default connection for four-gas mix, unless otherwise specified
•
at purchase.
Inlets 3-5 are designed connections for additional gases. However, unless
•
specified at the time of purchase, inlets 3-5 are configured for ambient air.
To confirm that the inlets are setup correctly, complete the following:
From the user options menu, press
1.
or
^ to scroll to the Inlet Setup
option.
Press
2.
sel to confirm the selection and access the inlet 1 (ambient air)
screen.
NOTe
Inlet 1 is designated for ambient air only and cannot be changed.
From the inlet
3.
1 screen, press
to access the inlet 2 screen.
Continue to press
to access the inlet 3, 4, and 5 screens. The O
2
gas type
displays when each of the inlet screens are accessed for the first time.
The inlet screens (2-5) have three options to modify:
Gas type
•
Gas concentration level
•
Gas cylinder lot # field
•
(field initially displays blank until data is entered).
Gas Type Screen