MicroDock II Automatic Test and Calibration Station
50
For information regarding solutions, refer to
Troubleshooting
.
NOTe
If a gas type fails, calibrate the detector prior to use.
Confirm that the gas concentration exceeds the low alarm setpoint of the
detector being tested (applicable to X5 only).
If an inlet is not setup correctly, the result of the bump check displays as the gas
type followed by a hyphen (–).
This example displays SO
2
, CO, LEL, and O
2
as gases not detected.
If a test gas fails, verify that the
gas cylinder is not empty,
•
connections are attached correctly,
•
inlets are setup correctly, and
•
gas concentration exceeds the low alarm setpoint (applicable to X5
•
detector).
Sensor Disabled
WArNINg
To prevent personal injury and/or property damage, replace the sensor
immediately. Refer to the corresponding detector user manual for sensor
replacement procedures.
If a sensor is disabled, results from the sensor do not display on the station LCD.
– displays where the gas type typically displays on the LCD.
NOTe
If a sensor is disabled, it is not recorded in the results log.