Note: A faster method of calibration is using the commands
[Measure frequency 0], [Measure frequency 1], [Measure frequency 2]
and [Measure frequency 3] commands instead of the balance test
command. The measured values are automatically entered in the
corresponding entities ([Frequency 0], [Frequency 1], [Frequency 2] and
[Frequency 3]). This method may not be followed when your gauge is
used for density measurement via the density displacer.
Entity name
Entity name
(SmartView)
Description
[Lock test]
[Lock test]
(Product level)
Raise the displacer as high as possible until it is in
full view, and good accessible.
[Freeze]
[Freeze]
(Product level)
Send a freeze command to stop the displacer
movement. Correct, if necessary temporarily the
motor limit high setting, in such a way that the
gauge will automatically return to this level after a
lock-test command.
[Set to
maintenance
mode]
[Maintenance mode]
(Product level)
Go into maintenance mode and do not enter any
command which is not specified below.
[Measure
frequency 0]
[Measure F0]
(Product level)
Measure the frequency of the force transducer with
a 25g weight.
When the measure frequency command is ready,
the status on the display changes MF to FR, the
displacer level primary value status code changes
from ‘Measure Frequency command active’ into
‘Freeze command active’ and the [Measured
frequency status] changes from ‘Command active’
to ‘Command not active’.
[Measure
frequency 1]
[Measure F1]
(Product level)
Measure the frequency of the force transducer with
a 100g weight.
[Measure
frequency 2]
[Measure F2]
(Product level)
Measure the frequency of the force transducer with
a 175g weight.
[Measure
frequency 3]
[Measure F3]
(Product level)
Measure the frequency of the force transducer with
a 250g weight.
[Measured
frequency status]
[MF Status]
‘Command active’ when the gauge is measuring
‘Command not active’ when the command is
finished or not started yet.
[Wire tension
protection]
[Wire tension pr]
(Product level)
Enable wire tension protection.
[Wire rupture
protection]
[Wire rupture pr]
(Product level)
Enable wire rupture protection.
[Set to
operational
mode]
[Operational mode]
(Product level)
Restart in operational mode
Summary of Contents for FlexLine SmartServo 954
Page 1: ...Service Manual 954 SmartServo FlexLine ...
Page 197: ...COMMISSIONING Part No 4417340 Revision 1 Service Manual 189 954 SmartServo FlexLine ...
Page 202: ...COMMISSIONING 194 Service Manual Part No 4417340 Revision 1 954 SmartServo FlexLine ...
Page 203: ...COMMISSIONING Part No 4417340 Revision 1 Service Manual 195 954 SmartServo FlexLine ...
Page 300: ...APPENDIX A 292 Service Manual Part No 4417340 Revision 1 954 SmartServo FlexLine ...
Page 302: ...APPENDIX B 294 Service Manual Part No 4417340 Revision 1 954 SmartServo FlexLine ...