COMMISSIONING
Part No.: 4417340 Revision 1
Service Manual
93
954 SmartServo FlexLine
7.3.4.16 Servo auto test
Name
Value Range
Default Value
Explanation
[SAT upwards]
With this command a
manual SAT upwards
can be performed.
[SAT downwards]
With this command a
manual SAT downwards
can be performed.
[Autonomous SAT enable] <SAT disable>
<SAT enable up>
<SAT enable up and
down>
<SAT enable up>
Determines whether the
SAT is disabled, or
enabled for upward
movements, or enabled
for upward and
downward movements.
[SAT raise height up]
<0.02>
Determines the upward
raise height.
[SAT raise height down]
<0.02>
Determines the
downward raise height.
[SAT interval time]
<60>
Determines how often
the SAT is performed.
[SAT sample interval]
<60>
Determines the time
period for checking if
there is a stable level.
[SAT trip distance]
<0.003>
Determines if the level
can be called stable.
That means if the
previous measurement
level compared to the
current measurement is
still within the trip
distance.
[SAT weight limit up]
<0.005>
Determines the
maximum weight
change if the displacer
is raised.
[SAT weight limit down]
<0.005>
Determines the
maximum weight
change if the displacer
is immersed.
[SAT immersed weight
limit]
<0.18>
Determines the
maximum total weight if
the displacer is
immersed.
Summary of Contents for FlexLine SmartServo 954
Page 1: ...Service Manual 954 SmartServo FlexLine ...
Page 197: ...COMMISSIONING Part No 4417340 Revision 1 Service Manual 189 954 SmartServo FlexLine ...
Page 202: ...COMMISSIONING 194 Service Manual Part No 4417340 Revision 1 954 SmartServo FlexLine ...
Page 203: ...COMMISSIONING Part No 4417340 Revision 1 Service Manual 195 954 SmartServo FlexLine ...
Page 300: ...APPENDIX A 292 Service Manual Part No 4417340 Revision 1 954 SmartServo FlexLine ...
Page 302: ...APPENDIX B 294 Service Manual Part No 4417340 Revision 1 954 SmartServo FlexLine ...