52
Overview
Settings operations procedure
Imports the master workpiece waveform as the master waveform, and sets the threshold values for
each judgment function from the master waveform.
Set the test conditions.
[OUTPUT]
Select the table to which to save
the master waveform.
[TABLE]
Enable the judgment conditions
to be used.
[JUDGE]
Calibrate the voltage and import the
master waveform.
[V CAL]
Imports the master data for the LC
and RC values and creates the PASS
judgment area.
(When using LC and RC values
judgments)
Set the judgment conditions.
[JUDGE]
Test conditions
settings mode
[SETTING]
Finish
1
Turns ON the
[ENABLE]
settings for the
judgment functions to be used. (p. 55)
•
[JUDGE]
> Use the judgment functions tab to turn ON
the
[ENABLE]
setting for the judgment functions to be
used, and to turn OFF the
[ENABLE]
setting for the
judgment functions that will not be used.
• Turning OFF the judgment functions that will not be
used shortens the test times.
2
Press
START
button to implement the impulse
tests and import the LC and RC master values
(Implemented multiple times). (p. 56)
• Up to 1024 LC and RC master values max. can be
imported.
•
Tap
[SCALE]
at the bottom of the screen to
automatically set the LC and RC graph scale values.
• Tap the master values on the LC and RC graph to
move the cursor.
•
Tap
[DELETE]
alongside the LC and RC graph to
delete the LC and RC master values selected by the
cursor.
•
Tap
[DELETE ALL]
alongside the LC and RC graph
to delete all the imported LC and RC master values.
3
The PASS judgment area is created
automatically from the imported LC and RC
•
Tap
[CREATE]
alongside the LC and RC graph>
Set the margin for the master values in the
PASS judgment area, and then tap
[CREATE]
to
automatically create the PASS judgment area.
• The PASS judgment area that has been created is
displayed in a gray square on the LC and RC graph.
4
Adjust the PASS judgment area manually.
• You can set the PASS judgment area manually using
the
[JUDGE]
>
[LCRC AREA]
tab.
You can set the PASS judgment area using the upper
and lower limit values of the LC and RC values when
[HI-LO]
is ON.
You can set the PASS judgment area using the LC
and RC values of the rectangle’s zenith when
[HI-LO]
is OFF.
• Tap the LC and RC values graph when the
[LCRC
AREA]
tab is selected to edit the PASS judgment
area directly.
5
Setting judgment conditions other than LC
and RC values judgments.
•
[JUDGE]
> Set the judgment threshold values in the
judgment functions tab.
Summary of Contents for ST4030
Page 2: ......
Page 14: ...6 Usage Notes ...
Page 32: ...24 Inspecting before Measurement and Verifying Operations ...
Page 42: ...34 Table Copying ...
Page 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Page 132: ...124 Terminal Open Error Setting ...
Page 198: ...190 Editing Files and Folders ...
Page 214: ...206 Interface Specifications ...
Page 248: ...240 Precautions when Processing Test Leads ...
Page 252: ...244 Index ...
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......