Waveform Judgments
Setting the differential surface area comparison judgment threshold value
(Measurement screen)
[MODE]
>
[SETTING]
>
[JUDGE]
>
[DIFF AREA]
2
1
3
1
Tap
[LIMIT]
.
2
Use the numeric keypad to enter the threshold values.
Settings range: 0.00% to 99.99% (resolution: 0.01%)
OFF
Differential surface area comparison value operations are implemented, but judgments are not.
3
Tap
[SET]
to confirm.
Setting the differential surface area comparison judgment area
(Measurement screen)
[MODE]
>
[SETTING]
>
[JUDGE]
>
[DIFF AREA]
/
Moves 1 point.
/
Moves 100 points.
Judgment area bar
[BEGIN]
: Start point
[END]
: End point
You can change the
judgment area by dragging
the lines of the start and
end points.
1
1
Set the start and end points for the judgment area.
5
Setting Judgment Conditions