197
Functional Specifications
Set judgment conditions
Waveform surface area
comparison judgment
[AREA]
Discharge judgments from the variation in the waveform surface area with the
same voltage.
Judgment values
Sets the deviation (
σ
) threshold value of the
waveform surface area
Settings range
OFF / 3 to 100
Discharge judgments
[DISCHARGE]
Discharge judgments from the variation in the discharge amount with the same
voltage.
Judgment values
Sets the deviation (
σ
) threshold value of the
discharge amount
Settings range
OFF / 3 to 100
LC and RC value judgments
[LCRC]
Discharge judgments from the variation in the LC and RC values with the same
voltage.
Judgment values
Sets the deviation (
σ
) threshold value of the LC
and RC values
Settings range
OFF / 3 to 100
Peak voltage value fluctuations
[Vpeak]
Discharge judgments from the variation in the peak voltage value with the same
voltage.
Judgment values
Sets the amount of variation in the first peak value
for each pulse relative to the center value for the
first peak values of all measurement pulses.
Settings range
OFF / 0% to 999%
Vibration frequency
fluctuations
[FREQ]
Discharge judgments from the variation in the vibration frequency with the same
voltage.
Judgment values
Sets the amount of variation in the vibration
frequency for each pulse relative to the
center value for the vibration frequency of all
measurement pulses.
Settings range
OFF / 0% to 999%
Safety protection function
Key lock function
Disables all settings changes and protects the settings details.
Enter the user-set pass code to unlock the settings.
Function settings
OFF/ON
Pass code
1 to 4-digit number
Interlock function
Enables test voltage output tests using external input signals.
Disables the interlocks by turning ON the external control terminal
(EXT. I/O) INTERLOCK
―
―
―
――
―
―――
pin when the setting is ON.
This interlock status disables testing for both key communications and external
control terminal (EXT. I/O).
Enter the user-set pass code to unlock the settings.
Function settings
OFF / ON
Pass code
1 to 4-digit number
Double action function
After the
STOP
button is pressed, the
START
button is enabled for only 1 sec.
Function settings
OFF / ON
13
Specifications
Summary of Contents for ST4030
Page 2: ......
Page 14: ...6 Usage Notes ...
Page 32: ...24 Inspecting before Measurement and Verifying Operations ...
Page 42: ...34 Table Copying ...
Page 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Page 132: ...124 Terminal Open Error Setting ...
Page 198: ...190 Editing Files and Folders ...
Page 214: ...206 Interface Specifications ...
Page 248: ...240 Precautions when Processing Test Leads ...
Page 252: ...244 Index ...
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......