44
Sampling Frequency/Number of Sampling Data
4.4
Sampling Frequency/Number of Sampling Data
Sets the voltage sampling frequency and the number of sampling data to be imported to the
instrument.
SAMPLING
(Sampling frequency)
Sets the voltage sampling frequency.
If a sufficiently long response waveform cannot be imported because
the response waveform vibration period is long, you can extend the
length of the waveform being imported by extending the sampling
frequency.
RECORD LENGTH
(No. of sampling data)
Sets the number of sampling data to be imported to the instrument.
Sets the length of the waveform to be imported, and sets the
sampling frequency. Adjust the number of sampling data so that a
suitable waveform length can be imported for judgment.
AUTO SET
(Automatic settings function for
range of waveform acquisition)
This function automatically adjusts and sets the sampling frequency
and number of sampling data during voltage calibration so that the
waveform acquisition range is optimal.
Sampling waveform lengths
Pulse application
area
Self-resonant area
Imported waveform length when the number of sampling data is 8001 points
Imported waveform length when the number of sampling data is
6001 points
Important: Recommended import length of the sampling waveform
Adjust the sampling wavelength so that the vibration waveform of the self-resonant area imports for 4 or
more periods to enable sufficient waveform data to be used by the LC and RC values operations and the
waveform judgment function. Slow the sampling frequency to lengthen the waveform to be imported.
Summary of Contents for ST4030
Page 2: ......
Page 14: ...6 Usage Notes ...
Page 32: ...24 Inspecting before Measurement and Verifying Operations ...
Page 42: ...34 Table Copying ...
Page 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Page 132: ...124 Terminal Open Error Setting ...
Page 198: ...190 Editing Files and Folders ...
Page 214: ...206 Interface Specifications ...
Page 248: ...240 Precautions when Processing Test Leads ...
Page 252: ...244 Index ...
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......