217
Error Display
WARNING
Display *
1
EXT.I/O
ERR pin
Communications
VCAL status
Meaning/Solution
WARNING 102
OFF
102
• Disturbances were detected in a number of
waveforms during voltage calibration.
• The calibration process may be experiencing
the effects of noise. Consider taking steps to
prevent noise.
• Probing may be unstable. Perform voltage
calibration after establishing stable probing.
• Check that the workpiece is normal.
• This error will result if the instrument
determines that this condition existed in 50%
or more of the response waveforms used in
calibration.
WARNING 103
OFF
103
• Discharges were detected in a number of
waveforms during voltage calibration.
• The calibration process may be experiencing
the effects of noise. Consider taking steps to
prevent noise.
• The set voltage may be too high. Lower the set
voltage and check instrument operation.
• This error will result if the instrument
determines that this condition existed in 50%
or more of the response waveforms used in
calibration.
WARNING 104
OFF
101
• The applied voltage failed to reach the set
voltage.
• This warning is displayed when TYPE 1
calibration is performed with the
[VOLT
ERROR]
setting disabled.
*1: If the voltage was calibrated from communications, the window is not displayed.
INFORMATION
Display *
1
EXT.I/O
ERR pin
Communications
VCAL status
Meaning
INFORMATION 1011
ON
1
Indicates that voltage calibration was canceled
due to a settings change.
INFORMATION 1012
ON
1
Indicates that voltage calibration was canceled
by the user.
14
Maintenance and Service
Summary of Contents for ST4030
Page 2: ......
Page 14: ...6 Usage Notes ...
Page 32: ...24 Inspecting before Measurement and Verifying Operations ...
Page 42: ...34 Table Copying ...
Page 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Page 132: ...124 Terminal Open Error Setting ...
Page 198: ...190 Editing Files and Folders ...
Page 214: ...206 Interface Specifications ...
Page 248: ...240 Precautions when Processing Test Leads ...
Page 252: ...244 Index ...
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......