47
Trigger Delay
4.5
Trigger Delay
Sets the delay time (trigger delay) from the start of measuring to the application of the first pulse.
If using this function, you can start measuring after the workpiece connection status has stabilized
even if trigger timing and probing timing are the same when combined in the equipment.
Trigger delay
Trigger
(Measurement start)
Pulse application
IMPORTANT
The
OUTPUT
lamp is synced to the EOM signals and so turns on even if the trigger is delayed.
4
Setting the
Test Conditions