LC and RC Values Area Judgments [LCRC AREA]
Importing LC and RC master values
Implement voltage calibration and import the master waveform before importing multiple LC and
RC values as the master values.
Test while implementing voltage calibration to save the LC and RC values acquired for each test as
master values to the instrument.
• Up to 1024 LC and RC master values max. can be imported.
• LC and RC master values are deleted when the power supply is turned OFF.
Checking and editing imported LC and RC master values
You can check imported LC and RC master values using the cursor. The cursor moves whenever a
LC or RC value is tapped.
•
Tap
[SCALE]
at the bottom of the screen to automatically set the LC and RC graph scale values.
•
Tap
[DELETE]
alongside the LC and RC graph to delete the LC and RC master values presently
selected by the cursor.
•
Tap
[DELETE ALL]
alongside the LC and RC graph to delete all the imported LC and RC master
values.