Functional Specifications
Break down voltage test (BDV) mode
Test overview
Tests the impulse while gradually increasing the voltage applied to the workpiece to
examine the discharge judgments.
Waveform surface area judgments, discharge judgments, LC and RC values area
judgments, peak voltage value fluctuations, and vibration frequency fluctuations are
used in discharge judgments.
Applied voltage settings
Sets the values of the start voltage, max. voltage, and voltage rise width
Start voltage
Settings range:
ST4030: 100 V to 3300 V (resolution: 10 V)
ST4030A: 100 V to 4200 V (resolution: 10 V)
Max. voltage
Settings range:
ST4030: 100 V to 3300 V (resolution: 10 V)
ST4030A: 100 V to 4200 V (resolution: 10 V)
Voltage rise width
Settings range:
ST4030: 10 V to 3200 V (resolution: 10 V)
ST4030A: 10 V to 4100 V (resolution: 10 V)
Maximum number of
steps
32 steps
Number of applied pulses
Sets the number of
application pulses and
the number of degaussing
pulses.
Degaussing pulse: Applied using the set number of
continuous pulses during BDV test startup
Application pulse: Applied using the set number of
continuous pulses for each step
Number of pulses: 3 to 32
Number of degaussing pulses: 0 to 10
Sets the pulse application
interval
Sets the minimum time of the pulse application interval
when the settings for the number of application pulses
and degaussing pulses is 2 or more.
Setting range: 0.050 s to 1.000 s (resolution: 0.001 s)
Sampling frequency
Sets the voltage sampling frequency
Setting range: 10 MHz, 20 MHz, 50 MHz, 100 MHz, 200 MHz
Number of sampling data
Sets the number of sampling data so that the required waveform length is imported.
Settings range: 1001 to 8001 points (1000 points step)
Summary of Contents for ST4030
Page 2: ......
Page 14: ...6 Usage Notes ...
Page 32: ...24 Inspecting before Measurement and Verifying Operations ...
Page 42: ...34 Table Copying ...
Page 108: ...100 Setting Dielectric Break down Judgment Conditions ...
Page 132: ...124 Terminal Open Error Setting ...
Page 198: ...190 Editing Files and Folders ...
Page 214: ...206 Interface Specifications ...
Page 248: ...240 Precautions when Processing Test Leads ...
Page 252: ...244 Index ...
Page 253: ......
Page 254: ......
Page 255: ......
Page 256: ......