Setting Dielectric Break down Judgment Conditions
Peak voltage value fluctuations
Discharges are detected based on variations in the first peak voltages among multiple pulses of
each applied voltage.
If the amount of variability in the first peak voltages for multiple pulses exceeds the set threshold,
the instrument will conclude that discharges are occurring.
(Measurement screen)
[MODE]
>
[BDV]
>
[JUDGE]
1
2
1
Tap
[Vpeak]
.
The numeric keypad window will open.
2
Set the deviation threshold value.
Setting range: 0% to 999%
3
Tap
[SET]
to confirm.