232
Max. Applicable Voltage (Reference Value) According to the Workpiece Inductance
15.6 Max. Applicable Voltage (Reference Value)
According to the Workpiece Inductance
The smaller the inductance of the workpiece being tested, the stronger the impact of voltage drop
inside the impulse winding tester, so the max. voltage that can be applied to the workpiece being
tested will become smaller.
If a greater voltage than the max. applicable voltage is set, an error will occur when “4.6 Voltage
Calibration” (p. 49) is implemented.
Describes the first peak voltage (reference value) when an impulse test is implemented on a low-
loss coil.
0
10
20
30
40
50
60
70
80
90
100
0
500
1000
1500
2000
2500
3000
3500
4000
4500
10
μ
H
100
μ
H
1 mH
10 mH
100 mH
First peak voltage [V]
First peak voltage when the discharge terminal
voltage is 4200 V
Workpiece inductance
Ratio
to
di
scharg
e
termi
nal
vol
tag
e
[%]