83
PW3335A985-01
Set and Query
:MEASure?
Output Items (Current Integration [Total Sum])
Syntax
:MEASure[:NORMal]:ITEM:IH:ALL
<Output item>
:MEASure[:NORMal]:ITEM:IH:CH1(?)
<Output item>
Data by current range during auto-range integration
:MEASure[:NORMal]:ITEM:IH:CH1:I200mA(?)
<Output item>
:MEASure[:NORMal]:ITEM:IH:CH1:I500mA(?)
<Output item>
:MEASure[:NORMal]:ITEM:IH:CH1:I1A(?)
<Output item>
:MEASure[:NORMal]:ITEM:IH:CH1:I2A(?)
<Output item>
:MEASure[:NORMal]:ITEM:IH:CH1:I5A(?)
<Output item>
:MEASure[:NORMal]:ITEM:IH:CH1:I10A(?)
<Output item>
:MEASure[:NORMal]:ITEM:IH:CH1:I20A(?)
<Output item>
:MEASure[:NORMal]:ITEM:IH:CH1:
BACKup(?)
<Output item>
Response
<Output item (NR1)>
128
64
32
16
8
4
2
1
bit7
bit6
bit5
bit4
bit3
bit2
bit1
bit0
DC
MN
ACDC
Description
Sets the current integration (total sum) data output items to a numerical value
between 0 and 11.
For example, specify 1 to output the AC+DC rectification or 2 to output the AC+DC
Umn rectification measurement value.
You can also output all rectification methods together at once. For example, you can
specify 3 to output both the AC+DC rectification and AC+DC Umn rectification
measurement values at the same time.
Although NRf numerical values are accepted, values to the right of the decimal are
dropped.
If no items are specified directly via a
:MEASure?
query, the measurement values
are output based on the specification of this command.
Example
Command
:MEAS:ITEM:IH:CH1 1
Specifies to output the instantaneous value of the AC/DC rectified current integration
(total sum).
Query
:MEAS:ITEM:IH:CH1?
Response
(When HEADER ON)
:MEASURE:NORMAL:ITEM:IH:CH1 1
(When HEADER OFF)
1
Note
・
If auto-range integration is ON, the output of the MN and ACDC will be invalid
data.
www.