56
PW3335A985-01
(8)
D/A output
Set and Query D/A Output (D/A1) Settings
Syntax
Command
Output Terminal D/A1
:AOUT
<Output item>
Query
Output Terminal D/A1
:AOUT?
Response
<Output item>
List of Measurement Item Specifications
<Output item> field.
Description
Command
D/A1 Output Item
Query
Returns the D/A1 output item.
(Same as
:AOUT:ITEM:DA1
.)
Example
Command
:AOUT S
Query
:AOUT?
Response
(When HEADER ON)
:AOUT:ITEM:DA1 S
(When HEADER OFF)
S
Note
You can use the AOUT:ITEM(?) command to query and set DA1 through DA7.
・
A device error will occur on units that do not have a D/A output.
Set and Query the D/A Output Terminal (D/A1 to D/A7) Output Items
Syntax
Command Output Terminal D/A1
:AOUT:ITEM:DA1
<D/A output item>
Output Terminal D/A2
:AOUT:ITEM:DA2
<D/A output item>
Output Terminal D/A3
:AOUT:ITEM:DA3
<D/A output item>
Output Terminal D/A4
:AOUT:ITEM:DA4
<D/A output item>
Output Terminal D/A5
:AOUT:ITEM:DA5
<D/A output item>
Output Terminal D/A6
:AOUT:ITEM:DA6
<D/A output item>
Output Terminal D/A7
:AOUT:ITEM:DA7
<D/A output item>
Query
Output Terminal D/A1
:AOUT:ITEM:DA1?
Output Terminal D/A2
:AOUT:ITEM:DA2?
Output Terminal D/A3
:AOUT:ITEM:DA3?
Output Terminal D/A4
:AOUT:ITEM:DA4?
Output Terminal D/A5
:AOUT:ITEM:DA5?
Output Terminal D/A6
:AOUT:ITEM:DA6?
Output Terminal D/A7
:AOUT:ITEM:DA7?
Response
<D/A output item>
See the
D/A output item specification list
for details.
Description
Sets or queries the output (rectification method) of the D/A output terminals
(DA1 to DA7).
Example
Command
:AOUT:ITEM:DA1 WP
Query
:AOUT:ITEM:DA1?
Response
(When HEADER ON)
:AOUT:ITEM:DA1 WP
(When HEADER OFF)
WP
Note
・
A device error will occur on units that do not have a D/A output.
www.
.com