33
PW3335A985-01
Standard Commands
(1)
System Data Command
Query Device ID (Identification Code)
Syntax
Query
*IDN?
Response
<Manufacturer name>, <Model name>, <Model type>, <Software version>,
<Serial number>
<Manufacturer name>
"HIOKI"
Fixed
<Model name>
"PW3335"
No. of channels: 1
<Model type>
RS-232c
GP-IB
D/A output
Current
Sensor
Input
00
●
-
-
-
01
-
●
-
-
02
●
-
●
-
03
●
-
-
●
04
●
●
●
●
Example
Query
*IDN?
Response
HIOKI,PW3335,04,V1.00,ser123456789
The Device ID is HIOKI PW3335-04 (RS-232c,GP-IB, with D/A
output,current sensor input), software version 1.00, 123456789.
Note
・
The response message has no header.
"*IDN?" must be the last query message in a program message.
・
Therefore, if any other query is detected after this query on the same line, a query error
will occur and no response message will be output.
Query Device Options
Syntax
Query
*OPT?
Response
<IF type>, <D/A output capability>,<
External current input sensor presence>
<IF type>
RS/GPIB/RS_GPIB/NONE
<D/A output capability>
DA_OUT/NONE
<
External current input
sensor presence>
CURR_SENSOR/NONE
Description
Queries the options available on the instrument.
Example
Query
*OPT?
Response
GPIB,DA_OUT,CURR_SENSOR
Instrument is equipped with a GP-IB interface, D/A output, and an external
current sensor.
Note
・
The response message has no header.
(2)
Internal Operation Command
Initialize Device
Syntax
Command
*RST
Description
Command
Resets the instrument to its initial state.
Note
・
Refer to the user's manual for the instrument (in the System Reset section)
for information about the initial settings.
・
The communications state is not initialized.
・
This command can be executed even when a system error has occurred.