21
PW3335A985-01
Device-specific Commands (Measurement Value Output)
Note:
:MEASure[:NORMAL]:ITEM:U:CH1(?)
→ Setting Command:MEASure[:NORMAL]:ITEM:U:CH1
Query :MEASure[:NORMAL]:ITEM:U:CH1?
Message ([ ]: Can be omitted)
Data Formats
(Response data
for queries)
Description
Reference
Page
:MEASure[:POWer]?
:MEASure[:NORMal]:VALue?
<Measurement Item 1> ...
Maximum 180
Queries measurement
data.
:MEASure:ITEM:ALLClear
Turns OFF all output items
(including harmonic wave).
:MEASure[:NORMal]:ITEM?
Queries output items.
:DATAout:ITEM(?)
(<Output Item 1>,
<Output Item 2>)
“:MEASure?” query output
specification
(3332-compatible)
:MEASure[:NORMAL]:ITEM:STATus:INST(?)
:MEASure[:NORMAL]:ITEM:STATus:MAXmin(?)
<
Output Item
0/1>
“:MEASure?” query
Set/Query
the measurement status
output.
:MEASure[:NORMAL]:ITEM:U:ALL
:MEASure[:NORMAL]:ITEM:U:CH1(?)
<Output Item
(Rectification Method)>
“:MEASure?” query
Sets/Queries
the voltage (instantaneous
value) data output.
:MEASure[:NORMAL]:ITEM:U_MAX:ALL
:MEASure[:NORMAL]:ITEM:U_MAX:CH1(?)
<Output Item
(Rectification Method)>
“:MEASure?” query
Sets/Queries the voltage
(maximum value) data
output.
:MEASure[:NORMAL]:ITEM:U_MIN:ALL
:MEASure[:NORMAL]:ITEM:U_MIN:CH1(?)
<Output Item
(Rectification Method)>
“:MEASure?” query
Sets/Queries the voltage
(minimum value) data
output.
:MEASure[:NORMAL]:ITEM:I:ALL
:MEASure[:NORMAL]:ITEM:I:CH1(?)
<Output Item
(Rectification Method)>
“:MEASure?” query
Sets/Queries the current
(instantaneous value) data
output.
:MEASure[:NORMAL]:ITEM:I_MAX:ALL
:MEASure[:NORMAL]:ITEM:I_MAX:CH1(?)
<Output Item
(Rectification Method)>
“:MEASure?” query
Sets/Queries the current
(maximum value) data
output.
:MEASure[:NORMAL]:ITEM:I_MIN:ALL
:MEASure[:NORMAL]:ITEM:I_MIN:CH1(?)
<Output Item
(Rectification Method)>
“:MEASure?” query
Sets/Queries the current
(minimum value)
data output.
:MEASure[:NORMAL]:ITEM:P:ALL
:MEASure[:NORMAL]:ITEM:P:CH1(?)
<Output Item
(Rectification Method)>
“:MEASure?” query
Sets/Queries the active
power (instantaneous
value) data output.
:MEASure[:NORMAL]:ITEM:P_MAX:ALL
:MEASure[:NORMAL]:ITEM:P_MAX:CH1(?)
<Output Item
(Rectification Method)>
“:MEASure?” query
Sets/Queries the active
power (maximum value)
data output.
:MEASure[:NORMAL]:ITEM:P_MIN:ALL
:MEASure[:NORMAL]:ITEM:P_MIN:CH1(?)
<Output Item
(Rectification Method)>
“:MEASure?” query
Sets/Queries the active
power (minimum value)
data output.
www.
.com
1.800.561.8187