PW3335A985-01
Set and Query Voltage Range Select
Syntax
Command
:VOLTage:SELect <data(NR1)>
Query
:VOLTage:SELect?
Response
<data>
128
64
32
16
8
4
2
1
<data(NR1)>
bit7
bit6
bit5
bit4
bit3
bit2
bit1
bit0
-
1000V
600V
300V
150V
60V
30V
15V
Description
Sets and queries voltage range use settings range key operation,
auto-range operation, and auto-range integration as a value from 0 to
127. Although NRf numerical values are accepted, values to the right of the
decimal are truncated.
Example
Command
:VOLT:SEL 124
Configures the setting so that the 15 V and 30 V ranges are skipped
.
Query
:VOLT:SEL?
Response
(When HEADER ON)
:VOLTAGE:SELECT 124
(When HEADER OFF)
124
Note
・
This command is provided to ensure compatibility with control programs for existing
models (3332). To take maximum advantage of the PW3335
’
s functionality, it is
recommended to use commands such as :VOLTage:SELect:6V.
・
This query cannot be used to query the 6 V range use setting.
・
If this command is used to set whether to select voltage ranges, the 6 V range use
setting will be set to OFF.
・
This command cannot be used to set the 1,000 V range (which is always ON).
・
Specifying a voltage range with the
“
:VOLTage:RANGe
”
command will cause range
select for the specified range to be turned on.
・
You cannot change this setting during integration, while the display is held, or when
the maximum/minimum values are being held.