84
PW3335A985-01
Set and Query
:MEASure?
Output Items (Positive Current Integration)
Syntax
:MEASure[:NORMal]:ITEM:PIH:ALL
<Output item>
:MEASure[:NORMal]:ITEM:PIH:CH1(?)
<Output item>
Data by current range during auto-range integration
:MEASure[:NORMal]:ITEM:PIH:CH1:I200mA(?)
<Output item>
:MEASure[:NORMal]:ITEM:PIH:CH1:I500mA(?)
<Output item>
:MEASure[:NORMal]:ITEM:PIH:CH1:I1A(?)
<Output item>
:MEASure[:NORMal]:ITEM:PIH:CH1:I2A(?)
<Output item>
:MEASure[:NORMal]:ITEM:PIH:CH1:I5A(?)
<Output item>
:MEASure[:NORMal]:ITEM:PIH:CH1:I10A(?)
<Output item>
:MEASure[:NORMal]:ITEM:PIH:CH1:I20A(?)
<Output item>
:MEASure[:NORMal]:ITEM:PIH:CH1:
BACKup(?)
<Output item>
Response
<Output item (NR1)>
128
64
32
16
8
4
2
1
bit7
bit6
bit5
bit4
bit3
bit2
bit1
bit0
DC
Description
Sets the positive current integration data output items to a numerical value between
0 and 8.
For example, specify 8 to output the DC rectification measurement value.
Although NRf numerical values are accepted, values to the right of the decimal are
dropped.
If no items are specified directly via a
:MEASure?
query, the measurement values
are output based on the specification of this command.
Example
Command
:MEAS:ITEM:PIH:CH1 8
Specifies to output the instantaneous value of the DC rectified positive current
integration.
Query
:MEAS:ITEM:PIH:CH1?
Response
(When HEADER ON)
:MEASURE:NORMAL:ITEM:PIH:CH1 8
(When HEADER OFF)
8
www.
.com
1.800.561.8187