66
PW3335A985-01
Toggle and Query the Normal Measurement and Harmonic Wave Measurement
Display Modes
Syntax
Command
:DISPlay:MODE
<NORM/HRMS/HCON
/HOSRMS/HOSCON>
Query
:DISPlay:MODE?
Response
<NORM/HRMS/HCON/HOSRMS/HOSCON>
NORM: Normal measurement values
HRMS: Harmonic wave level display, all orders
(a: order, b/c/d: harmonic wave measurement)
HCON: Harmonic wave content display, all orders
(a: order, b/c/d: harmonic wave measurement)
HOSRMS: Harmonic wave level display, individual orders
(a/b/c/d: harmonic wave measurement)
HOSCON: Harmonic wave content display, individual orders
(a/b/c/d: harmonic wave measurement)
Description
Toggles or queries the content of the display area (normal measurement or
harmonic wave measurement).
Example
Command
:DISP:MODE NORM
Query
:DISP:MODE?
Response
(When HEADER ON)
:DISPLAY:MODE NORM
(When HEADER OFF)
NORM
Set and Query the Displayed Order for Harmonic Wave Common Order Display Mode
Syntax
Command
:DISPlay:HARMonic:ORDer
<0 to 50 (NR1)>
Query
:DISPlay:HARMonic:ORDer?
Response
<0 to 50 (NR1)>
Description
Sets or queries the order for harmonic wave common order display mode.
Although NRf numerical values are accepted, values to the right of the
decimal are dropped.
Example
Command
:DISP:HARM:ORD 21
Query
:DISP:HARM:ORD?
Response
(When HEADER ON)
:DISPLAY:HARMONIC:ORDER 21
(When HEADER OFF)
21
Note
・
If the display is in any other mode other than harmonic wave (all orders) display mode,
the display will not be immediately affected by this command.
The setting will be applied when the display mode is changed via a command such
as
:DISPlay:MODE
.
Set and Query the Display Items for Harmonic Wave Common Order Display Mode
Syntax
Command
Display Area: b
:DISPlay:HARMonic:B:ITEM
<
Harmonic wave display item
>
Display Area: c
:DISPlay:HARMonic:C:ITEM
<
Harmonic wave display item
>
Display Area: d
:DISPlay:HARMonic:D:ITEM
<
Harmonic wave display item
>
Query
Display Area: b
:DISPlay:HARMonic:B:ITEM?
Display Area: c
:DISPlay:HARMonic:C:ITEM?
Display Area: d
:DISPlay:HARMonic:D:ITEM?
Response
<Harmonic wave display item>
Harmonic wave voltage
HU (HU1
can also be used.
)
Harmonic wave current
HI (HI1
can also be used.
)
the harmonic wave
active power output
HP (HP1
can also be used.
)
Description
Sets or queries the display items for harmonic wave common order display
mode.
Example
Command
:DISP:HARM:B:ITEM HU
Query
:DISP:HARM:B:ITEM?
Response
(When HEADER ON)
:DISPLAY:HARMONIC:B:ITEM HU
(When HEADER OFF)
HU
Note
・
If the display is in any other mode other than harmonic wave (all orders) display mode,
the display will not be immediately affected by this command.
Change the display mode via a command such as
:DISPlay:MODE
.
www.
.com
1.800.561.8187