104
PW3335A985-01
Set and Query
:MEASure:HARMonic?
Output Items (Instantaneous, Maximum, and
Minimum Values for the Harmonic Wave Voltage Phase Angle)
Syntax
Instantaneous
value
:MEASure:HARMonic:ITEM:UPHAse:ALL <Output item>
:MEASure:HARMonic:ITEM:UPHAse:CH1(?) <Output item>
Maximum
value
:MEASure:HARMonic:ITEM:UPHAse_MAX:ALL <Output item>
:MEASure:HARMonic:ITEM:UPHAse_MAX:CH1(?) <Output item>
Minimum value
:MEASure:HARMonic:ITEM:UPHAse_MIN:ALL <Output item>
:MEASure:HARMonic:ITEM:UPHAse_MIN:CH1(?) <Output item>
Response
<Output item (NR1)>
128
64
32
16
8
4
2
1
bit7
bit6
bit5
bit4
bit3
bit2
bit1
bit0
HUPHAse
Description
Sets the harmonic wave voltage phase angle data output items returned
by
:MEASure:HARMonic?
(instantaneous value, maximum value, minimum value)
as numerical values between 0 and 1.
The order output is the order specified via
MEASure:HARMonic:ITEM:ORDer
.
Although NRf numerical values are accepted, values to the right of the decimal are
dropped.
Example
Command
:MEAS:HARM:ITEM:UPHA:CH1 1
Specifies to output the instantaneous value of the harmonic wave voltage phase
angle.
Query
:MEAS:HARM:ITEM:UPHA:CH1?
Response
(When HEADER ON)
:MEASURE:HARMONIC:ITEM:UPHA:CH1 1
(When HEADER OFF)
1
Note
・
This command is used along with the
:MEASure:HARMonic:ITEM:ORDer
command to
specify the harmonic wave output order.
・
If the specification is the same as that for the
:MEASure:HARMonic:ITEM:LIST
command,
the command issued afterwards takes precedence.
・
Harmonic wave phase angle data cannot be displayed on the instrument. This data can only
be obtained through communications commands.
Set and Query
:MEASure:HARMonic?
Output Items (Instantaneous, Maximum, and
Minimum Values for the Harmonic Wave Current Phase Angle)
Syntax
Instantaneous
value
:MEASure:HARMonic:ITEM:IPHAse:ALL <Output item>
:MEASure:HARMonic:ITEM:IPHAse:CH1(?) <Output item>
Maximum
value
:MEASure:HARMonic:ITEM:IPHAse_MAX:ALL <Output item>
:MEASure:HARMonic:ITEM:IPHAse_MAX:CH1(?) <Output item>
Minimum
value
:MEASure:HARMonic:ITEM:IPHAse_MIN:ALL <Output item>
:MEASure:HARMonic:ITEM:IPHAse_MIN:CH1(?) <Output item>
Response
<Output item (NR1)>
128
64
32
16
8
4
2
1
bit7
bit6
bit5
bit4
bit3
bit2
bit1
bit0
HIPHAse
Description
Sets the harmonic wave current phase angle data output items returned
by
:MEASure:HARMonic?
(instantaneous value, maximum value, minimum value)
as numerical values between 0 and 1.
The order output is the order specified via
MEASure:HARMonic:ITEM:ORDer
.
Although NRf numerical values are accepted, values to the right of the decimal are
dropped.
Example
Command
:MEAS:HARM:ITEM:IPHA:CH1 1
Specifies to output the instantaneous value of the harmonic wave current phase
angle.
Query
:MEAS:HARM:ITEM:IPHA:CH1?
Response
(When HEADER ON)
:MEASURE:HARMONIC:ITEM:IPHASE:CH1 1
(When HEADER OFF)
1
Note
・
This command is used along with the
:MEASure:HARMonic:ITEM:ORDer
command
to specify the harmonic wave output order.
・
If the specification is the same as that for the
:MEASure:HARMonic:ITEM:LIST
command,
the command issued afterwards takes precedence.
・
Harmonic wave phase angle data cannot be displayed on the instrument. This data can
only be obtained through communications commands.
www.
.com