57
PW3335A985-01
Set and Query Current Range Select during D/A Output Terminal (D/A 1 to 7)
Integrated Value Output
Syntax
Command Output Terminal D/A1
:AOUT:IRANge:DA1
<
integrated value current range
(NR2)
>
Output Terminal D/A2
:AOUT:IRANge:DA2
<
integrated value current range
(NR2)
>
Output Terminal D/A3
:AOUT:IRANge:DA3
<
integrated value current range
(NR2)
>
Output Terminal D/A4
:AOUT:IRANge:DA4
<
integrated value current range
(NR2)
>
Output Terminal D/A5
:AOUT:IRANge:DA5
<
integrated value current range
(NR2)
>
Output Terminal D/A6
:AOUT:IRANge:DA6
<
integrated value current range
(NR2)
>
Output Terminal D/A7
:AOUT:IRANge:DA7
<
integrated value current range
(NR2)
>
Query
Output Terminal D/A1
:AOUT:IRANge:DA1?
Output Terminal D/A2
:AOUT:IRANge:DA2?
Output Terminal D/A3
:AOUT:IRANge:DA3?
Output Terminal D/A4
:AOUT:IRANge:DA4?
Output Terminal D/A5
:AOUT:IRANge:DA5?
Output Terminal D/A6
:AOUT:IRANge:DA6?
Output Terminal D/A7
:AOUT:IRANge:DA7?
Response
<
integrated value current range
(NR2)>
0
0.2 / 0.5 / 1 / 2 / 5 / 10 / 20
BACKUP
Description
Integrated values exist for each current range when auto-range integration
is performed. This command sets or queries the current range for which to
output data when outputting integrated values. (Unit: ampere [A])
NRf numerical values are accepted.
When the value 0 is selected, integrated values for all ranges are added,
and the resulting total value is output.
Example
Command
Returns the current range setting as an NR2-format value or string.
Query
:AOUT:IRANge:DA1?
Response
(When HEADER ON)
:AOUT:IRANGE:DA1 0.2
(When HEADER OFF)
0.2
Note
・
A device error will occur on units that do not have a D/A output.
・
When auto-range integration is off, changing the setting results in an execution error. For DA
output, the total value will be output regardless of the setting.
・
Specifying a value other than <integrated value current range (NR2)> will cause the setting
to be changed to
“
range that can measure the specified value.
”
However, the next range up
will be selected if the range
’
s full-scale value is exceeded.
・
If a negative value is specified, the absolute value will be used.
www.
.com
1.800.561.8187