76
PW3335A985-01
Set and Query
:MEASure?
Output Items
Syntax
Command
:DATAout:ITEM
<data1 (NR1)>,<data2 (NR1)>
Query
:DATAout:ITEM?
Response
<data1>,<data2>
128
64
32
16
8
4
2
1
<data1 (NR1)>
bit7
bit6
bit5
bit4
bit3
bit2
bit1
bit0
FREQU
DEG
PF
Q
S
P
I
U
<data2 (NR1)>
bit7
bit6
bit5
bit4
bit3
bit2
bit1
bit0
IPK
IH
TIME
MWP
PWP
WP
Description
Sets or queries the measurement items for the :MEASure? query as a
numerical value between 0 and 255.
Although NRf numerical values are accepted, values to the right of the
decimal are dropped.
Example
Command
:DATA:ITEM 7,0
(This enables the output of the voltage, current, and active power.)
Query
:DATA:ITEM?
Response
(When HEADER ON)
:DATAOUT:ITEM 7,0
(When HEADER OFF)
7,0
Note
・
This command is provided for compatibility with control programs for the
existing model 3332.
This command can be used only to set or query AC/DC rectification measurement
values.
In order to fully utilize the capabilities of the PW3335, we recommend using the
MEASure:ITEM
commands.