9.4 Alarm Signal Output (Alarm Output)
162
This signal is output when alarm criteria are satisfied. Specify the desired alarm criteria.
9.4 Alarm Signal Output (Alarm Output)
Alarm signal output
(ALARM)
When alarm criteria are satisfied
About the Alarm Output Signal
Output signal
Open-drain output (with voltage output), active LOW
Output voltage range
HIGH level: 4.0 to 5.0 V, LOW level: 0 to 0.5 V
Output Refresh
Every recording interval
Maximum input voltage
30 V DC, 5 mA max, 200 mW max
GND
ALARM
10 k
Ω
5 V
For the connection
method (p. 158)
Connect the ALARM (No. 4) and GND (No. 1) terminals of the
external control terminal block to the signal output line and
ground, respectively. When the alarm criteria are satisfied, a
Low level (0 to 0.5V) signal is output.
The ALARM pin implements active-low operation. In the example connection, current flows to the
coil when there is no alarm output. Choose a relay with a contact architecture that implements the
desired operation.
Relay
FET, etc.
GND
GND
ALARM
10 k
Ω
5 V DC
Instrument
DC power supply
(Provide a DC power supply that accom-
modates the relay coil’s rated voltage.)
ALARM pin circuit architecture and example relay connection
Summary of Contents for LR8432-20
Page 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Page 2: ......
Page 22: ...Operating Precautions 14...
Page 50: ...2 5 Zero Adjustment 42...
Page 88: ...3 7 Viewing All Channel Settings in a List 80...
Page 104: ...4 5 About Recording Operation 96...
Page 154: ...7 4 Waveform Calculation Formula 146...
Page 164: ...8 3 Making System Settings 156...
Page 232: ...Appendix 12 Heat Flow A44...
Page 236: ...Index Index 4...
Page 237: ......
Page 238: ......