7.1 Performing Numerical Calculation
135
3
7
C
h
ap
te
r 7
N
u
m
e
ri
cal
/W
avef
orm
C
a
lc
u
lat
ions
4
Split Calc. is
set to [On].
4
Configure the split calculation setting. (Select
[Close]
. at the end.)
Setting options: (
∗
default setting)
Off
∗
Only the calculation value obtained last is saved.
On
Calculation values obtained from the measurement start during
regular time intervals (
[Length]
) are saved. Set
[Length]
to the
time interval for obtaining calculation values.
Ref Time
The first interval is automatically adjusted so that the next calcula-
tion value obtained during regular time interval that starts at the
time set in Split Time box is saved. (Only the first interval becomes
shorter than the interval set in
[Split Length]
.) Set the reference
time and time interval in
[Split Time]
and
[Split Length]
, respec-
tively.
3
4
4
Split Calc. is set
to [Ref Time].
Summary of Contents for LR8432-20
Page 1: ...EN LR8432 20 HEAT FLOW LOGGER Instruction Manual Jan 2016 Edition 1 LR8432B980 00 16 01H...
Page 2: ......
Page 22: ...Operating Precautions 14...
Page 50: ...2 5 Zero Adjustment 42...
Page 88: ...3 7 Viewing All Channel Settings in a List 80...
Page 104: ...4 5 About Recording Operation 96...
Page 154: ...7 4 Waveform Calculation Formula 146...
Page 164: ...8 3 Making System Settings 156...
Page 232: ...Appendix 12 Heat Flow A44...
Page 236: ...Index Index 4...
Page 237: ......
Page 238: ......